Inspecting a Surface

Three optical and probe tools at the MPaCT Lab in Flagstaff, Arizona, look at a surface. They answer different questions. A colour photograph is not a roughness number, and a millimetre map is not a 50 µm AFM scan.

Start here: do you need a picture, a millimetre-scale height map, or a nanometre height on a small patch?

Decision table

Digital microscopy Optical profilometry AFM
Output Colour image, on-screen 2D measure Calibrated 3D height, Ra, Rq, step height Calibrated nanometre height
Field Set by the objective Up to 50 × 50 mm 50 × 50 µm
Contact None None Physical tip
Chemistry None None None
Instrument at MPaCT Keyence VHX-7000 Keyence VK-X3000 AFM Workshop B-2

SEM still owns morphology plus composition. This page is for the three tools people confuse with each other before they reach vacuum.

Choose by the question

"What does this defect look like in colour?" - VHX-7000.

"What is Ra over a millimetre?" - VK-X3000.

"What is the height of this 40 µm patch in nanometres?" - AFM.

"Is the film continuous, and what is it made of?" - SEM, not these three.

A full microscopy decision that includes TEM is TEM vs SEM vs AFM.

Frequently asked questions

Should I use a digital microscope, an optical profiler, or AFM?

Use the VHX-7000 when you need a colour image and an on-screen measurement. Use the VK-X3000 when you need calibrated roughness or step height over millimetres. Use the AFM Workshop B-2 when you need a nanometre height map on a 50 by 50 micrometre patch. They are not substitutes.

Can a digital microscope replace AFM for roughness?

No. A focus-stacked image looks three-dimensional. It is not a calibrated height map. Ra and Rq over millimetres belong on the VK-X3000. Nanometre roughness on a small, flat patch belongs on the B-2.